MassBank Record: EA030301



 Caffeine; LC-ESI-ITFT; MS2; CE: 35%; R=7500; [M+H]+ 
Mass Spectrum
Chemical Structure
Generated by the Chemistry Development Kit (http://github.com/cdk) \n

ACCESSION: EA030301
RECORD_TITLE: Caffeine; LC-ESI-ITFT; MS2; CE: 35%; R=7500; [M+H]+
DATE: 2014.01.14
AUTHORS: Stravs MDepartment of Environmental Chemistry, Eawag, Schymanski EDepartment of Environmental Chemistry, Eawag, Singer HDepartment of Environmental Chemistry, Eawag, Department of Environmental Chemistry, Eawag
LICENSE: CC BY
COPYRIGHT: Copyright (C) 2012 Eawag, Duebendorf, Switzerland
COMMENT: CONFIDENCE standard compound
COMMENT: EAWAG_UCHEM_ID 303

CH$NAME: Caffeine CH$NAME: 1,3,7-trimethyl-3,7-dihydro-1H-purine-2,6-dione CH$NAME: 1,3,7-trimethylpurine-2,6-dione CH$COMPOUND_CLASS: N/A; Environmental Standard CH$FORMULA: C8H10N4O2 CH$EXACT_MASS: 194.0804 CH$SMILES: c12c(n(c(=O)n(c1=O)C)C)ncn2C CH$IUPAC: InChI=1S/C8H10N4O2/c1-10-4-9-6-5(10)7(13)12(3)8(14)11(6)2/h4H,1-3H3 CH$LINK: CAS 58-08-2 CH$LINK: CHEBI 27732 CH$LINK: HMDB HMDB01847 CH$LINK: KEGG C07481 CH$LINK: PUBCHEM CID:2519 CH$LINK: INCHIKEY RYYVLZVUVIJVGH-UHFFFAOYSA-N CH$LINK: CHEMSPIDER 2424 CH$LINK: COMPTOX DTXSID0020232
AC$INSTRUMENT: LTQ Orbitrap XL Thermo Scientific AC$INSTRUMENT_TYPE: LC-ESI-ITFT AC$MASS_SPECTROMETRY: MS_TYPE MS2 AC$MASS_SPECTROMETRY: ION_MODE POSITIVE AC$MASS_SPECTROMETRY: IONIZATION ESI AC$MASS_SPECTROMETRY: FRAGMENTATION_MODE CID AC$MASS_SPECTROMETRY: COLLISION_ENERGY 35 % (nominal) AC$MASS_SPECTROMETRY: RESOLUTION 7500 AC$CHROMATOGRAPHY: COLUMN_NAME XBridge C18 3.5um, 2.1x50mm, Waters AC$CHROMATOGRAPHY: FLOW_GRADIENT 90/10 at 0 min, 50/50 at 4 min, 5/95 at 17 min, 5/95 at 25 min, 90/10 at 25.1 min, 90/10 at 30 min AC$CHROMATOGRAPHY: FLOW_RATE 200 ul/min AC$CHROMATOGRAPHY: RETENTION_TIME 4.0 min AC$CHROMATOGRAPHY: SOLVENT A water with 0.1% formic acid AC$CHROMATOGRAPHY: SOLVENT B methanol with 0.1% formic acid
MS$FOCUSED_ION: BASE_PEAK 195.088 MS$FOCUSED_ION: PRECURSOR_M/Z 195.0877 MS$FOCUSED_ION: PRECURSOR_TYPE [M+H]+ MS$DATA_PROCESSING: DEPROFILE Spline MS$DATA_PROCESSING: RECALIBRATE loess on assigned fragments and MS1 MS$DATA_PROCESSING: REANALYZE Peaks with additional N2/O included MS$DATA_PROCESSING: WHOLE RMassBank 1.3.1
PK$SPLASH: splash10-000i-0900000000-695d910d49fc0beb1d54 PK$ANNOTATION: m/z tentative_formula formula_count mass error(ppm) 110.0713 C5H8N3+ 1 110.0713 0.24 138.0662 C6H8N3O+ 1 138.0662 -0.06 195.0877 C8H11N4O2+ 1 195.0877 0.14 PK$NUM_PEAK: 3 PK$PEAK: m/z int. rel.int. 110.0713 33560 21 138.0662 1570359.5 999 195.0877 238647.1 151 //

MassBank | Copyright Line
Copyright © since 2006 MassBank Project
Copyright © since 2011 NORMAN Association
Copyright © since 2017 MassBank Consortium
Responsible: Dr. Tobias Schulze (tobias.schulze@ufz.de)