MassBank Record: MSJ00090



 Benzene; CI-Q; MS; Positive; Reactant ion NO+ 
Mass Spectrum
Chemical Structure
Generated by the Chemistry Development Kit (http://github.com/cdk) \n

ACCESSION: MSJ00090
RECORD_TITLE: Benzene; CI-Q; MS; Positive; Reactant ion NO+
DATE: 2018.12.19
AUTHORS: , Satoshi Inomata, Hiroyuki Yamada, Hiroshi Tanimoto, Center for Environmental Measurement and Analysis and Center for Global Environmental Research, National Institute for Environmental Studies, and Department of Mechanical Engineering, Tokyo Denki University
LICENSE: CC BY
COPYRIGHT: Satoshi Inomata, Hiroyuki Yamada, Hiroshi Tanimoto, Center for Environmental Measurement and Analysis and Center for Global Environmental Research, National Institute for Environmental Studies, and Department of Mechanical Engineering, Tokyo Denki University.
PUBLICATION: Hiroyuki Yamada, Satoshi Inomata, Hiroshi Tanimoto, Atmos. Environ., 107, 166-173 (2015).
COMMENT: The detection sensitivity of benzene was determined by data at 0, 20, 40, 60, 80, 100 ppb(v/v). A ten-VOC premixed standard gas containing methanol, acetonitrile, acetaldehyde, acetone, isoprene, benzene, methylvinylketone, toluene, p-xylene, and 1,3,5-trimethylbenzene was used for the determination of the detection sensitivity.
COMMENT: The reactant ion, NO+, was generated from the air at the hollow cathode discharge 2.0 mA by using Proton Transfer Reaction plus Switchable Reagent Ion Mass Spectrometry, PTR+SRI-MS.
COMMENT: Pressure inside of the drift tube was 2.1 mbar. Temperature of the drift tube was 105 C. Electric field of the drift tube was E/N = 67 Td (U_drift = 250 V). Temperature of inlet tube was 105 C.
COMMENT: This record was created by the financial support of MEXT/JSPS KAKENHI Grant Number 16HP2005 to the Mass Spectrometry Society of Japan.

CH$NAME: Benzene CH$COMPOUND_CLASS: Synthetic compound; aromatic compound CH$FORMULA: C6H6 CH$EXACT_MASS: 78.0470 CH$SMILES: c1ccccc1 CH$IUPAC: InChI=1S/C6H6/c1-2-4-6-5-3-1/h1-6H CH$LINK: CHEMSPIDER 236 CH$LINK: INCHIKEY UHOVQNZJYSORNB-UHFFFAOYSA-N CH$LINK: PUBCHEM CID:241 CH$LINK: COMPTOX DTXSID3039242
AC$INSTRUMENT: PTR-QMS 500 (IONICON Analytik GmbH, Innsbruck, Austria) AC$INSTRUMENT_TYPE: CI-Q AC$MASS_SPECTROMETRY: MS_TYPE MS AC$MASS_SPECTROMETRY: ION_MODE POSITIVE AC$MASS_SPECTROMETRY: IONIZATION CI AC$MASS_SPECTROMETRY: REACTANT_ION NO+
PK$SPLASH: splash10-004i-9200000000-d4888ed94a6a4e783628 PK$NUM_PEAK: 2 PK$PEAK: m/z int. rel.int. 78 5.2 999 108 1.7 327 //

MassBank | Copyright Line
Copyright © since 2006 MassBank Project
Copyright © since 2011 NORMAN Association
Copyright © since 2017 MassBank Consortium
Responsible: Dr. Tobias Schulze (tobias.schulze@ufz.de)